Identifying electrode bridging from electrical distance distributions: A survey of publicly-available EEG data using a new method

Daniel M. Alschuler1, Craig E. Tenke1,2, Gerard E. Bruder1,2, Jürgen Kayser1,2

1Division of Cognitive Neuroscience, New York State Psychiatric Institute, New York, NY, USA; 2Department of Psychiatry, Columbia University College of Physicians and Surgeons, New York, NY, USA

Received 27 February 2013; revised 23 July 2013; accepted 25 August 2013; available online 1 October 2013; published 3 March 2014.

Abstract

Objective. EEG topographies may be distorted by electrode bridges, typically caused by electrolyte spreading between adjacent electrodes. We therefore sought to determine the prevalence of electrode bridging and its potential impact on the EEG literature.
Methods. Five publicly-available EEG datasets were evaluated for evidence of bridging using a new screening method that employs the temporal variance of pairwise difference waveforms (electrical distance). Distinctive characteristics of electrical distance frequency distributions were used to develop an algorithm to identify electrode bridges in datasets with different montages (22-64 channels) and noise properties.
Results. The extent of bridging varied substantially across datasets: 54% of EEG recording sessions contained an electrode bridge, and the mean percentage of bridged electrodes in a montage was as high as 18% in one of the datasets. Furthermore, over 40% of the recording channels were bridged in 9 of 203 sessions. These findings were independently validated by visual inspection.
Conclusions. The new algorithm conveniently, efficiently, and reliably identified electrode bridges across different datasets and recording conditions. Electrode bridging may constitute a substantial problem for some datasets.
Significance. Given the extent of the electrode bridging across datasets, this problem may be more widespread than commonly thought. However, when used as an automatic screening routine, the new algorithm will prevent pitfalls stemming from unrecognized electrode bridges.

doi:10.1016/j.clinph.2013.08.024

Key Words: electroencephalogram (EEG); electrolyte bridge; artifact; electrical distance; intrinsic Hjorth; public dataset